Simultaneous determining multiple trace proteins in urine with surface plasmon resonance biosensor
Objective To establish a novel methodology for simultaneous quantification of four trace proteins in urine with surface plasmon resonance (SPR) biosensor. Methods Monoclonal antibodies against Alpha-1 microglobulin (A1M)
Beta-2 microglobulin (B2M)
microalbumin (MA) and immunoglobulin G (IgG) were immobilized on the surface of biosensor chips
and then SPR was adopted to detect the corresponding antigens in standard solution containing the above-mentioned proteins. Calibrations of the 4 proteins were performed and the methodological evaluations we
re conducted to validate the sensitivity and specificity. Clinical urine samples from 125 inpatients were collected
and the contents of these 4 proteins were detected with SPR and immuno-turbidimetric assay at the same time. The results from the 2 methods were compared. Results The proposed SPR immunosensor had good signal-noise ratio and specificity. All the R
2
of the calibration curves were higher than 0.98. The sensitivity of A1M was 5 μg/L
B2M 7.3 μg/L
MA 11.5 μg/L
and IgG 22 μg/L. All the detection was finished in 20 min. This proposed method had good consistency when compared with traditional immuno-turbidimetric assay. Conclusion This proposed SPR immunosensor chip can accurately and simultaneously determine multiple trace proteins in urine
and can be a useful alternative to current methodology.