AI Jianguo,Chu Tongwei,Zhou Yue,Chen Weijun.Design and biomechanical test of a novel lumbosacral anterior plate[J].J Amry Med Univ (J Third Mil Med Univ),2010,32(08):826-829.
AI Jianguo,Chu Tongwei,Zhou Yue,Chen Weijun.Design and biomechanical test of a novel lumbosacral anterior plate[J].J Amry Med Univ (J Third Mil Med Univ),2010,32(08):826-829.DOI:
Design and biomechanical test of a novel lumbosacral anterior plate
Objective To design a new lumbosacral anterior plate by statistical analysis for the numerical value of the anatomic structure in Chinese. Methods The numerical value of lumbosacral angle and height of lumbosacral vertebrae
and discus intervertebralis were measured in 350 orthotropia and lateral view Xray pictures of lumbosacral vertebrae of adult normal individuals. According to the acquired amplitude of lumbosacral vertebrae in Chinese
we designed the modified lumbosacral anterior p
late
and then tested the plate in 4 different states of working condition on 8 fresh spinal columns to assay biomechanical data
and the results were compared with those of lumbosacral anterior secure plate (PACH). Results The amplitude of lumbosacral angle were 125 to 135
the heights of anterior border L
5
and S
1
vertebrae were (27.06±2.30) and (22.3±2.8)mm
the height of L
5
S
1
discus intervertebralis was (8.2±0.3)mm. The flex bias and the compression rigidity had significant deviation for the vertebral column fixed the modified plate [flexation (4.706±0.012)mm
extension (4.549±0.298)mm
lateral bending (5.412±0.384)mm
average value of rigidity in 400 N (105.8±7.6)N/mm
]
and PACH[flexation (5.241±0.113) mm
extension (5.662±0.246)mm
lateral bending (6.767±0.017)mm
and average value of rigidity (87.4±4.2)N/mm
]
respectively
with statistical significance among them (P0.05). Conclusion The modified lumbosacral anterior plate system accommodates the anatomic structure of lumbosacral vertebrae in Chinese
possesses conspicuous stability
and is suitable for reconstruction of lumbosacral vertebrae in indicated patients.
One-stage anterior internal fixation with multidirectional self-locking plate in treatment of lumbosacral tuberculosis: report of 35 cases
Biomechanical comparison of 4 plate internal fixation methods for simulated fractures of acetabular posterior column
Finite element simulation and biomechanical analysis of indirect injury of temporomandibular joint under lateral impact on the angle of human mandible
Clinical outcomes of distal tibia posterior locking plate internal fixation for posterior Pilon fractures with lateral malleolus fracture through a combined posterolateral approach
Threedimensional finite element simulation and biomechanical analysis of maxillofacial blast injuries and secondary injuries to the skull base