GUAN Xiao,ZHANG Hui-jing,BI Ying-nan,ZHANG Li,HAO Dun-ling,WEI Jin-yu.A microfluidic chip to concentrate and capture EHEC O157∶H7[J].J Amry Med Univ (J Third Mil Med Univ),2009,31(10):887-890.
GUAN Xiao,ZHANG Hui-jing,BI Ying-nan,ZHANG Li,HAO Dun-ling,WEI Jin-yu.A microfluidic chip to concentrate and capture EHEC O157∶H7[J].J Amry Med Univ (J Third Mil Med Univ),2009,31(10):887-890.DOI:
A microfluidic chip to concentrate and capture EHEC O157∶H7
Objective To study the efficiency of concentration and capture of enterohemorrhagic Escherichia coli (EHEC) O157∶H7 by a self-made microfluidic chip based on microbead. Methods The anti-EHEC O157∶H7 antibody was first attached on the microbeads by direct covalent linkage
then the microbeads were filled into microfluidic chip. Thus
the bacteria capture chip was set up. Totally 1.6×10
9
CFU fluorescence labeled bacteria flowed into the chip at a sampling speed of 1
2 and 5 μl/min respectively
and the elution speed of 0.5
2
5
10 and 20 μl/min. The specific capture efficiency of the chip was strengthened through the optimization of sampling speed
elution speed and time. Results The best sampling speed
elution speed and time were 5
μl/min
10 μl/min and 5 min respectively. There was significant difference between the test group and the control group (P0.01). Conclusion Our self-made bacteria capture chip is of high catching efficiency
which lay a good base for the construction of rapid detection microfluidic system of bacteria.