YANG Fan,HUANG Hui,PU Xiao-yun.Rapid detection of TSH and T4 by microfluidic chip-based immunoassays[J].J Amry Med Univ (J Third Mil Med Univ),2007,29(23):2254-2256.
YANG Fan,HUANG Hui,PU Xiao-yun.Rapid detection of TSH and T4 by microfluidic chip-based immunoassays[J].J Amry Med Univ (J Third Mil Med Univ),2007,29(23):2254-2256.DOI:
Rapid detection of TSH and T4 by microfluidic chip-based immunoassays
Objective To establish an immunoassay based on microfluidic chip for detection of TSH and total T
4
. Methods The direct-write laser micromachining technology was used to fabricate microfluidic chip
luminol-H
2
O
2
as luminator. This method and Roche E2010 electrochemiluminescence immunoassay were used to examine TSH by double-antibody sandwich method and total T
4
by competitive principle in 50 serum samples collected at our outpatient clinic. Results The detection was completed within 25 min. Ten microliter sample and reagent was required. The detection range of TSH was 0-50 μIU/ml and its average intraassay and interassay
precision was 4.7% and 4.6% respectively and its average recovery rate was 95.3%. The detection range of T
4
was 9.375-300 ng/ml
and its average intraassay and interassay precision was 4.9% and 4.7% respectively and its average recovery rate was 91.27%. Conclusion Rapid detection of TSH and T
4
by microfluidic chip-based immunoassays is convenient
stable and accurate and of less reagent and time consuming.